At the end of the Switch Operation Menu is the Loop Check function. Using the MODE key
to select the Loop Check function. After lop is displayed for 1 second, the value of the lat-
est measurement mode is displayed as default. Use keys to set displayed value any-
where between –1999 and 1999 to confirm switch operation and/or analog output operation.
#!$#"
);1+3-A78-:)<176;
In all setting modes, values are set with the keys. Use the key to increase and
the key to decrease the value. A repeat state occurs in three phases of speed when the
keys are pressed for more than 0.5 seconds to increase or decrease numerical value.
keys are also used for setting switch, unit and filter in the function setting mode.
!-(-:77."-6;7:
In measurement mode, select display to adjust zero using key. Press MODE + keys for
more than 3 seconds (unit “---” display blinks) after releasing pressure from the pressure port.
Automatic zero adjustment takes place approximately 1 second later for pressure display to
be zero.
When the zero adjustment is successful, Adj is displayed.
Error displays e-0 for 1 second when applied pressure was outside of range of –10 to 10%
FS, zero adjustment will not be allowed.